Contained within the IP Active layer, are the ____ _____ crystals.
Barium Fluorohalide
15
(T or F) The CR latent image is considered to be 'trapped' electrons.
True
15
What causes the electrons to be released from the F centers, within the phosphors of the CR IP ?
laser
15
The latent image will lose about ____% of its energy over about 8 hours time.
25%
15
(T or F) CR systems are considered to be "Stimulated Phosphorescence"
True
15
The movement of the laser across the imaging plate (vs the plate moving), is called:
Fast Scan
15
(T or F) To correct the elliptical shape of the laser, the beam is “shaped” by special optics that keep the beam size, shape, and speed largely independent of the beam position.
True
15
The reader scans the plate with red light in a zigzag, or _______ pattern.
raster
15
The letters, PM as part of a PM tube stands for:
Photomultiplier
15
This device converts analog data into digital data.
ADC (Analog to Digital Converter)
15
To digitize the analog signal from the PMT, it must first be _______.
Sampled
15
Increasing the sampling frequency results in a smaller sampling and pixel pitch, which (increases or decreases?) the spatial resolution of the digital image.
increases
15
The distance between the midpoint of one pixel to the midpoint of an adjacent pixel describes the _____ _______.
pixel pitch
15
An IP is 'erased' via exposure to what color light ?
intense white light
15
IP’s are very sensitive to scatter and should be erased how often? to prevent the buildup of any background signal (or if unsure of last use)